A Comprehensive Calibration Strategy for Texas Department of Transportation Falling Weight Deflectometer Fleet
-
2003-11-01
-
Details
-
Creators:
-
Corporate Creators:
-
Corporate Contributors:
-
Subject/TRT Terms:
-
Publication/ Report Number:
-
Resource Type:
-
Geographical Coverage:
-
Edition:Interim Report: July 2001 -July, 2002
-
Corporate Publisher:
-
Abstract:The degree of accuracy with which the Falling Weight Deflectometer (FWD) data are collected and analyzed has a direct impact on the conclusions drawn in many aspects of TxDOT operation. Since TxDOT owns 15 FWD units, one of the largest FWD fleets of any single transportation agency in the world, high levels of repeatability and reproducibility are considered to be of major importance for an adequate interchangeability of the FWD fleet. As summarized in Research Report 1784-1, the existing calibration protocol improves the repeatability of each FWD but it does not address the reproducibility of the fleet. In this report a new calibration process has been recommended that seems to improve the reproducibility of the fleet. The calibration process consists of three steps: a routine and diagnostic maintenance, a simple calibration procedure followed by a comprehensive one if necessary.
-
Format:
-
Funding:
-
Collection(s):
-
Main Document Checksum:urn:sha-512:7251056641743d64ac26dc2671a8300a7c953c40ff581c098ff5d406a15551b71244823246f57c9f540bc61b84b30d94670523ad2677aa7aa49c4923fef7beff
-
Download URL:
-
File Type: