Abstract:In an earlier study under the contract grant number BDV25-977-06, two portable Fourier transform infrared spectrometer
(FTIR) were evaluated for their ability to quickly detect and/or quantify the presence of polymer modifiers, including styrene
butadiene styrene (SBS) polymer and/or ground tire rubber (GTR), in asphalt binder. This report summarizes a follow-up
study that evaluates a third brand of portable FTIR device, the Agilent Handheld ExoScan 4300 FTIR Spectrometer, for
identification and/or quantification of polymer modifiers in asphalt binder.
The same binder samples used in the earlier study were tested with the Agilent Handheld ExoScan 4300 FTIR Spectrometer.
The test data were analyzed and compared with the results from the other FTIR devices included in the earlier study. It was
found that the Agilent Handheld ExoScan 4300 FTIR Spectrometer can capture absorbance spectra similar to those from
other portable FTIR devices and can read absorbance spectrum from asphalt mixture samples. The variability in test data of
the Agilent Handheld ExoScan 4300 FTIR Spectrometer, however, is higher than that of the TruDefender FTX FTIR
spectrometer.
-